![]() |
|
ABSTRACT Gas analysis close to the boiler exit often gives the
most valuable process control information but poses serious problems
for the measuring instrumentation. The gas typically will be at high
temperature, will be strongly stratified and may contain high burdens
of particulate material. The normal choice to be made is between extractive
and in situ gas monitors but both have disadvantages in this
type of application. Extractive systems Use conventional gas analyzers
in conjunction with sophisticated sampling systems which clean and dry
the gas. These can lead to unacceptably long response times and unpredictable
changes in sample composition. More importantly, sampling systems comprising
filters, pumps and bubblers require frequent and expensive maintenance.
In situ analyzers, although more robust, do not match up to conventional
analyzers in terms of accuracy and reliability when operating in such
harsh conditions. This paper describes a hybrid of the two types of systems
utilized to measure the concentration of carbon monoxide (CO) at the
economizer outlet of a gas-fired utility boiler. Samples are extracted
from the gas duct using multiple probes which are strategically located
to ensure a sample composition which is representative of the bulk stream.
This sample is cooled, passed through a simple water ex-tractor and
analyzed by an in situ gas monitor without further treatment.
Instead of an inductor or vacuum pump, the sample is extracted by using
the negative pressure of the forced-draft fan. Results are presented from a trial application on a
Louisiana power station and their use and interpretation are discussed. INTRODUCTION On-line measurement of the constituents of flue gas
has for a long time been troublesome for the utility industry. This
paper addresses some of those problems and a new solution for measuring
carbon monoxide (CO). We have combined the best features from both the
ex-traction type and the in situ type for this solution. The
goal is to have a measurement which can be used for closed loop control
and be acceptable for environmental reporting. Because of the slow speed
of response, stack mounted methods are not addressed. EXTRACTION TYPE The extraction method removes a sample from a vessel
and delivers it to a measurement instrument - usually in a different
location. This method has served the industry for a number of years
even though there was considerable maintenance expense on the system.
A variety of measurement -methods were used including chemical cells,
thermal conductivity, thermo-magnetic and infrared. Most of these methods
worked well when the sample was delivered clean and at a dew point lower
than the operating temperature of the measuring device. The sample conditioning is the main problem with extraction
systems. To clean up the sample, filters are used on the sample probes
and throughout the sample system. Filters add time delays to the sample
stream. Other sample conditioning methods are necessary, such as liquid
bubbler chambers, flow and pressure regulators, all of which add volume
to the sample system. Added volume adds delay to the sample stream.
Delays of 30 to 120 seconds are common. Time constants of this magnitude
added to the inherent delays of the flue gas path, make closed loop
control very difficult and of questionable value. Moving the sample point as near as possible to the
furnace exit will reduce the overall delay but it creates a problem
with gas sample temperature. The higher sample temperature requires
greater cooling capacity and larger volume bubblers for sample conditioning.
It is essential to maintain the sample dew point below the analyzer
operating temperature. Allowing condensate to form in the analyzer will
quickly damage the analyzer cells. These systems are notorious for their
high maintenance cost. The best feature of extraction systems is the flexibility
of averaging multiple sample points for a truly representative sample. IN SITU
TYPE In situ type analyzers
measure the flue gas constituent in place. For CO measurement some type
of infrared analyzer is generally used. In situ analyzers have
good response time and work well in closed loop control of CO. The most
widely used location of infrared in situ CO analyzers is across
a duct carrying flue gas. For control, the preferred location is at
the economizer outlet provided the gas temperature will be below 700'F.
Above 700'F, the accuracy of these instruments falls off rapidly. There is another problem with across the duct in
situ measurement due to horizontal and vertical stratification.
Stratification in a duct has many widely varying patterns. To obtain
a representative sample would require several sighting paths each requiring
a complete infrared analyzer system. Obviously this would be very costly. The sensitivity and accuracy of the infrared in
situ analyzer has proven to be sufficient for tile required environmental
documentation when it can see a representative sample. EXTRACTION / IN SITU
METHOD "HEUSZOMETER" The Heuszometer consists of a multi-probe sample extraction
system, a simplified -moisture removal device and a measurement chamber
using an infrared in situ type CO analyzer. The probe assembly
is designed to extract a sample from a cross section of the duct at
the economizer outlet. Each probe is an assembly of 3 to 5 sample tubes.
A combination of probe assemblies and tube lengths is strategically
placed to give a good representative sample. Each of the sample tubes
in one duct is connected to the moisture removal device. Care is taken
to maintain equal flow in each tube by maintaining equal resistance
in each flow path. The tubing is run in such a way as to obtain some
sample cooling before reaching the moisture removal device. This device
serves several functions, including mixing the multiple sample streams
and cooling the sample. The internal design consists of cone-shaped
discs placed in the path of the incoming samples so that proper mixing
of the samples occurs. At the same time, additional sample cooling is
accomplished by chilling the dispersion cones using an air operated
energy conversion device. This chamber cools the sample below its dew
point and the moisture is dropped out of the sample and discharged to
a drain. The composite sample is sent to a measuring chamber.
An infrared CO analyzer is mounted on the measuring chamber. The design
of the measuring chamber is such that ample sensitivity is obtained
to comply with environmental requirements and to provide sufficiently
rapid response for use in closed loop control. The exhaust line from
the measuring chamber connects to a point on the unit with a lower pressure
than the point of the sample. At Gulf States Utilities, we installed
the exhaust line to vent into the intake of the forced draft fans. By
this method, we have been able to have response times as short as seven
seconds and with average times of 10 to 12 seconds. These response times
are well within the realm of good closed loop control. RESULTS OF TRIAL APPLICATION The Heuszometer combines the best of two worlds in
the use of its sample probes, which have been proven in the extraction
systems and the in situ measurement which works well when it
can see a representative sample. The Heuszometer Mark II system was
installed in 1982 at Nelson Station in southwestern Louisiana. The system
has worked well since that time and has provided much information for
operating the unit as well as for designing changes incorporated into
the Mark M. The Mark M version has just been installed on one unit at
Sabine Station. Plans are to install this system on five additional
units in Texas. The Heuszometer Mark U provided operator guidance for
the set point on 02 automatic control from 1982 through 1991 with a
minimum of maintenance. When the analog analyzer began needing maintenance,
it was replaced with a newer digital version. During the nine year trial
period, operations occasionally questioned the CO reading because the
O., reading at the economizer outlet would be higher than the textbook
version of CO versus 0.2. The operators became believers only after
several verifications of the readings were made by our performance department's
test van. High temperature oxygen probes were installed in the furnace
which eliminated the penthouse leakage in the 02 measurement and this
convinced all doubters that the CO measurement was correct. Since the utility industry must comply with new environmental
standards, a reliable CO measurement will be essential. To reduce NOX,
the logical approach is to do those things which have the least cost,
like low excess air firing, burners out of service, individual burner
flame shaping techniques, and other attempts to change the temperature
profile in the furnace. Under several of these conditions, stoichiometric
or even sub-stoichiometric may be required. When this is the situation,
a fast, reliable CO measurement is required. The Heuszometer Mark HI was recently tested at Sabine
Station. Its accuracy was verified by standard calibration gases as
well as in actual operation on the unit at 400 megawatts (MW). The system
was operated in parallel with the performance test van and it became
a question as to which system was the standard. The Mark III had about
a 10-second response time which was 40 seconds faster than the test
van. CONCLUSION As utility companies tighten up the firing of their-
units and get reliable CO measurements, the inadequate wind box distribution
and unreliable burner air registers will become apparent and troublesome.
One technique which the Heuszometer uses is to take single sample tube
readings so that a duct profile of CO is possible. The proper use of
this method will be a big assist in adjusting burners. The Mark D[I Heuszometer is now being built by the
analyzer supplier. This new system should fill the need for CO measurement
and control on most fossil fuel fired utility boilers. Our testing has
been done on natural gas fired boilers. We see little, if any, changes
needed for use on oil fired units. A patent has been applied for both
the hardware and the process. Plans are being made under a separate project to do
testing on coal fired units for possibly using filters on the probes
or increasing the tube size of each probe. The Heuszometer is a proven
design for oil and gas. When further testing is done on coal fired units
a paper will be submitted on that outcome. In the meantime, the 'Heuszometer'
is available for gas or oil fired units. |